In brief
- In the standards, “EDOF” is not the name of an optical mechanism but a label assigned according to clinical outcome against a monofocal lens: depth of focus greater by at least 0.5 D, better intermediate vision and non-inferior distance vision (AAO 2017, ANSI Z80.35-2018).
- ISO 11979-7:2024 divides simultaneous vision lenses into three groups (MIOL, EDF, FVR), all of them defined in a controlled trial against a monofocal lens. It has no category for monofocal plus lenses.
- In 2024 the ESCRS working group proposed a functional classification based on the defocus curve: PARTIAL-RoF (narrow, enhance, extend) and FULL-RoF (continuous, smooth, steep).
- Dr. Stern and Prof. Gatinel argued that a defocus curve measures the whole eye and not the lens, and proposed classifying by means of individualized optical modeling.
- Different mechanisms can fall into the same functional category, and each mechanism has its own optical price. It is worth knowing both.
Three languages that get mixed up
The ESCRS working group distinguishes three categories that must not be mixed: lens characteristics (diffractive grating, refractive zone, small aperture, material, platform), optical outcome (light distribution, MTF, PSF) and functional outcome (visual acuity at different distances, defocus curve, spectacle independence, dysphotopsia). In their words, assigning a lens to a category such as “trifocal” or “EDOF” is to this day influenced more by expert opinion and manufacturers' claims than by research.
In the review they published in 2025, Dr. Stern and Prof. Gatinel add examples of the confusion: some manufacturers call a lens “EDOF” to distance it from the halo image of multifocal lenses; a lens classified as monofocal plus in one region is labeled EDOF in another; and certain diffractive lenses are described as “low add multifocal,” a term they recommend not using.
The AAO and ANSI criteria
In 2017 an AAO task force, working with FDA staff, published a clinical definition of an EDOF lens, and it was adopted in the American standard ANSI Z80.35-2018. A lens is considered an EDOF only if it meets all the conditions, in a trial against a monofocal control group:
| Condition | Requirement |
|---|---|
| Monocular depth of focus | At least 0.5 D greater than that of the control, at a threshold of 0.2 logMAR (20/32) |
| Intermediate vision (DCIVA, 66 cm, photopic) | Statistical superiority over the control, and at least half of the eyes (the median) at 0.2 logMAR or better |
| Corrected distance vision (BCDVA) | Non-inferiority to the control, with a margin of 0.1 logMAR |
| Study size (AAO) | At least 100 patients in the EDOF group |
The task force also defined how to measure: a monocular defocus curve with best distance correction, from +1.50 to −2.50 D in steps of 0.5 D (and of 0.25 D around zero), with randomized letters. The “depth of focus” is the range from zero diopters to the most negative value at which mean acuity is 0.2 logMAR or better. Important for our purposes: the guidance requires measuring the pupil on the day of testing and stratifying the results by pupil size (below 3, 3–4 and above 4 mm) and by axial length. In other words, it was already clear then that the curve depends on the eye and not only on the lens. According to the ESCRS, ANSI also requires a monotonic decrease in acuity from distance to near, and if the curve has an inflection point, it must not exceed 0.04 logMAR.
ISO 11979-7:2024
The international standard defines four main categories: monofocal, toric, simultaneous vision lenses (SVL) and accommodating lenses. The SVLs are divided into three:
- MIOL: distance and near (and with trifocal lenses, intermediate as well) better than with a monofocal lens.
- EDF: distance through intermediate. ISO agrees with ANSI on all the conditions and adds an absolute range of at least 1.5 D at the 0.2 logMAR threshold.
- FVR (full visual range): distance, intermediate and near, with a range of at least 2.5 D.
The ESCRS working group lists three limitations: every assignment requires a randomized trial against a monofocal lens, which has not been conducted for most models; the categories are not mutually exclusive (an FVR lens also meets the MIOL conditions); and there is no category for monofocal plus lenses.
And what about monofocal plus?
A 2023 scoping review tested enhanced monofocal lenses against the ANSI conditions: in no study did they reach an added 0.5 D of depth of focus, and no study showed a median DCIVA of 0.2 logMAR. The authors' conclusion: until a suitable standard exists, it is correct to classify them as monofocal. A 2025 meta-analysis of defocus curves (36 studies) arrived at a similar picture: a mean depth of focus of 1.17 D with monofocal lenses, 1.28 D with monofocal plus and 1.81 D with EDOF lenses, with no significant difference between the first two at any point on the curve. Some of the authors are affiliated with a lens manufacturer, which also funded the writing assistance.

The ESCRS functional classification (2024)
The classification was built from a cluster analysis of published monocular defocus curves: 107 studies were included in the review, of which 69 reports with a monocular curve entered the analysis. Two metrics were enough to separate the groups: the range of field (RoF) at a threshold of 0.2 or 0.3 logMAR, and the improvement in acuity from intermediate to near (DVA) when the curve does not decrease monotonically.
| Category | RoF at 0.2 logMAR | RoF at 0.3 logMAR | DVA | Conventional equivalent |
|---|---|---|---|---|
| PARTIAL – narrow | < 1.2 D | < 1.61 D | 0 | Monofocal |
| PARTIAL – enhance | 1.2–1.58 D | 1.61–1.98 D | 0 | Monofocal plus |
| PARTIAL – extend | 1.58–2.3 D | 1.98–2.75 D | 0 | EDOF |
| FULL – continuous | ≥ 2.3 D | ≥ 2.75 D | < 0.05 | Trifocal or bifocal with a low add |
| FULL – smooth | ≥ 2.3 D | ≥ 2.75 D | 0.05–0.14 | Trifocal with a high add |
| FULL – steep | ≥ 2.3 D | ≥ 2.75 D | ≥ 0.14 | Bifocal with a high add |

The authors themselves note the main limitation: the classification describes only acuity across the range, and not dysphotopsia or patient-reported outcomes, because reporting of these in studies is not uniform. They also explain why they did not classify by optical bench data: there are no agreed thresholds for the through-focus response.
The Stern and Gatinel critique
Already in the 2025 review, Dr. Stern and Prof. Gatinel wrote that a good classification should also include the optical mechanism, because it has clinical significance: diffractive lenses are associated with more halos, and refractive lenses are more pupil-dependent. In 2026 they published a letter in J Cataract Refract Surg that focuses on the method itself:
- The defocus curve is a measure of the system. It combines the cornea (about two-thirds of the refractive power, with wide variability in aberrations), the anterior chamber depth and the axial length (which change the lens position and the vergence geometry), the pupil, and the tilt and decentration of the lens.
- The measurement is noisy. Visual acuity varies even in the same eye under the same conditions; defocus testing is long and tiring; the samples are small, the examiners many, and the data sometimes retrospective.
- The mean hides the spread. At every point on the curve the standard deviation is large, but what is published is a smoothed mean, which may present a range that no single eye achieves.
- The aberrations of the cornea and the lens add up. A cornea with positive SA can partially offset a lens with negative SA and narrow the range, while in a neutral cornea the same lens will give a wider range and lower contrast. A mean of such eyes reflects the population, not the lens.
Their proposal: classify by individualized optical modeling (ray tracing or wavefront simulation) that incorporates the measured topography, anterior chamber depth, axial length and centration, under standardized pupil conditions. The working group published a reply in the same issue. On the interaction between the SA of the cornea and that of the lens, see spherical aberration: a clinical guide.
Disclosure: Prof. Gatinel is a consultant to BVI. Dr. Stern is a co-founder and shareholder of AIOLsci, which develops personalized simulation of lens performance.
The optical principle and what it does to the curve
The functional classification tells you what the patient sees on average; the optical principle tells you what the price is and in which eyes the lens may behave differently.
| Principle | How it extends the focus | Typical behavior on the curve | Common category | The main price |
|---|---|---|---|---|
| Diffractive multifocal | A kinoform grating splits the light into diffraction orders; the step height determines the light distribution and the step width the add | Separate peaks on the bench; on the acuity curve, a dip at intermediate (bifocal) or a more continuous curve (trifocal) | FULL (steep, smooth or continuous) | Halos and glare; a slight loss of contrast; sensitivity to residual refractive error |
| Diffractive EDOF | Few, wide steps and a low add; a design that offsets part of the chromatic aberration | Extended monotonic decrease | PARTIAL – extend | Halos without the near benefit of a multifocal lens |
| Refractive EDOF (a strong central zone or negative SA) | The power increases toward the center | A broad, low peak and a monotonic decrease; varies with the pupil | PARTIAL – extend | Dependence on the pupil and on corneal SA; “daytime myopia” with a small pupil |
| Monofocal plus | A subtle change in central power, or higher-order asphericity | A peak slightly wider than that of a monofocal lens | PARTIAL – enhance or narrow | Modest benefit; pupil dependence in some models |
| Small aperture (pinhole) | Blocks the peripheral rays and reduces the blur circle | A broad monotonic decrease | Partial correction | Less light reaching the retina; poor vision in dim light |
| Spiral | Refractive structures of different powers that wind around one another | A continuous range; in the lens that was tested, mainly thanks to symmetric SA | There are not yet enough clinical data | New technology, under evaluation |
Two notes on the table. First, this is typical behavior and not a law: the category of a given lens is determined by its own data. Second, the boundaries are blurring: new diffractive lenses with many peaks give a continuous curve and remain multifocal, and a refractive lens with many rings can come close to a multifocal. With the spiral lens, for example, wavefront decomposition in the laboratory showed that the spiral component contributes little, and that the extended range (about 2.5–3.0 D in simulation) stems mainly from symmetric SA. See the optical characterization of the spiral lens, small-aperture IOLs and monofocal plus IOLs.
Take-home messages
- When a manufacturer writes “EDOF,” check what it rests on: a controlled trial according to ANSI or ISO, or an uncontrolled case series.
- With a defocus curve, ask: monocular or binocular, with distance correction, at what pupil size, how many eyes, and where are the standard deviations.
- The functional category does not predict dysphotopsia; the optical mechanism gives a better clue.
- A mean curve is not the eye in front of you: the cornea, the pupil and the centration change it. See pupil size and corneal SA.
- Monofocal plus is best presented to the patient as a monofocal lens with a modest intermediate boost, not as an EDOF.
Frequently asked questions
Is a monofocal plus lens an EDOF lens?
According to the AAO and ANSI criteria, no. In a 2023 review, no study showed that these lenses add 0.5 D to the depth of focus or reach a median intermediate acuity of 0.2 logMAR, and the authors recommended classifying them as monofocal. In the ESCRS functional classification they have a category of their own: PARTIAL-RoF enhance.
Can a lens be classified from the manufacturer's through-focus graph?
Not formally. The ESCRS working group notes that there are no agreed thresholds for the through-focus response on the optical bench, so a classification based on it often rests on personal impression. The graph is useful for understanding the optical principle, but it describes one model eye, one aperture and usually one wavelength.
References
- Ribeiro F, Dick HB, Kohnen T, Findl O, Nuijts R, Cochener B, Fernández J. Evidence-based functional classification of simultaneous vision intraocular lenses: seeking a global consensus by the ESCRS Functional Vision Working Group. J Cataract Refract Surg. 2024;50(8):794-798. Link
- Stern B, Gatinel D. Comment on: Evidence-based functional classification of simultaneous vision intraocular lenses: seeking a global consensus by the ESCRS Functional Vision Working Group. J Cataract Refract Surg. 2026;52(5):523. Link
- MacRae S, Holladay JT, Glasser A, et al. Special Report: American Academy of Ophthalmology Task Force Consensus Statement for Extended Depth of Focus Intraocular Lenses. Ophthalmology. 2017;124(1):139-141. Link
- Fernández J, Ribeiro F, Rocha-de-Lossada C, Rodríguez-Vallejo M. Functional classification of intraocular lenses based on defocus curves: a scoping review and cluster analysis. J Refract Surg. 2024;40(2):e108-e116. Link
- Stern B, Gatinel D. Presbyopia correction in lens replacement surgery: a review. Clin Exp Ophthalmol. 2025;53(6):668-681. Link
- Fernández J, Rocha-de-Lossada C, Zamorano-Martín F, Rodríguez-Calvo-de-Mora M, Rodríguez-Vallejo M. Positioning of enhanced monofocal intraocular lenses between conventional monofocal and extended depth of focus lenses: a scoping review. BMC Ophthalmol. 2023;23:101. Link
- Srinivasan S, Nyankerh C, Hull J, Suryakumar R. Meta-analysis of defocus curves of monofocal, enhanced monofocal and extended depth of focus IOLs. BMJ Open Ophthalmol. 2025;10(1):e002025. Link
- Gatinel D, Entin A, Stern B. In vitro optical characterization of the RayOne Galaxy spiral extended depth-of-focus intraocular lens using high-resolution Mach–Zehnder interferometry. Biomed Opt Express. 2026;17(7):3984-4001. Link