In brief
- Asking “which multifocal lens is the best” is like asking “which IOL power is the best”: the answer depends on the eye.
- The workflow: cornea (topography, SA, irregularity) → pupil (before dilation) → lens design → simulation in complex eyes.
- The cornea can itself be “multifocal” or “EDOF,” and corneal aberrations can cancel those of the lens or amplify them.
- Sometimes a “premium” result can be achieved even with a monofocal lens, once the optics of the cornea and the pupil are understood.
How not to choose a multifocal lens
- Attending a manufacturer-sponsored session.
- Asking a colleague whether their new lens is “good.”
- Choosing the cheapest lens.
- Implanting only the model that is available at your hospital.
None of these approaches takes the patient's eye into account. The market is competitive, models change quickly, and marketing terms such as “wavefront stretching” or “pure refractive optics” are no substitute for understanding the optical profile.
The analogy with lens power
Before Fyodorov introduced modern IOL power calculation in 1967, it was common practice to implant a lens of a single standard power and correct the residual error with glasses. No one would do that today: from SRK to PEARL-DGS, biometry formulas match the power to the eye. Choosing the model is even more complex than choosing the power, and yet in most cases it is still done without a model of the eye.
Step 1: The cornea
- Keratometry samples only the center (2.3–2.5 mm, and with the Lenstar also 1.65 mm). Choosing the model requires the shape of the entire cornea.
- Topography is mandatory in a premium practice, and not only for toric lenses: it detects irregularity (coma) and assesses asphericity and SA.
- Q is not SA: at K = 44 D, an average cornea (Q ≈ −0.26) leaves about +0.19 µm; only Q ≈ −0.54 cancels the SA. The population mean is +0.28 µm (range 0.055–0.544, 6 mm pupil); after myopic ablation the mean is +0.36 µm, with values up to +1.02 µm, and after hyperopic ablation the mean is close to zero, with a range from −0.58 to +0.63 µm, depending on the amount of correction.
- Myth: “After LASIK, EDOF is preferable only because the power calculation is less accurate.” Sometimes that is true and sometimes it is not: the cornea itself may behave like an EDOF or a multifocal lens, and corneal aberrations can cancel those of the lens.
Combinations after LASIK
| Cornea | IOL | Expected effect |
|---|---|---|
| After myopic LASIK (with high positive SA) | LuxSmart (with negative SA) | Mutual cancellation, partial or complete, similar to a monofocal lens in a normal cornea |
| After myopic LASIK | Monofocal | “Free” EDOF, similar to the RayOne EMV |
| After myopic LASIK | RayOne EMV (with positive SA) | Stronger EDOF, but the add is in the periphery, working against the near reflex |
| After hyperopic LASIK (with negative SA) | Monofocal | An Eyhance-like effect |
| After hyperopic LASIK | Eyhance | Amplification of the EDOF effect |
Positive SA and negative SA both extend the focus. Negative SA is usually more comfortable, because the add is in the center, and the pupil constricts with the near reflex.
A case from the clinic
A cornea with asymmetric inferior steepening: along the 90° meridian, keratometry rises from 45.8 D superiorly to 50.9 D inferiorly. Implanting a toric monofocal lens in such an eye gives, under photopic conditions, a depth of focus similar to that of a patient with a normal cornea and a Lentis Mplus Toric lens. In one case the depth of focus comes from corneal coma, and in the other from coma created by the lens, but the optical effect is similar. If the cornea already provides the effect “for free,” it can be put to use.
Dr. Stern's position on asymmetric segmental refractive lenses (such as the Lentis Mplus): coma degrades image quality more than SA does, and the concept, borrowed from bifocal glasses, does not work the same way in the eye, which cannot “move” between the zones of the lens. Such lenses may cause image shift and dysphotopsia more than halos. That said, many surgeons report excellent clinical results.
Step 2: The pupil
- Measure it before dilation. It is one of the most important parameters, and few surgeons measure it routinely.
- Cataract patients have small pupils: outdoors, 2.39 mm on average (1.44–3.03); senile miosis of about 1.72 mm between the ages of 20 and 60 in dim light; and after surgery the pupil becomes smaller still, by 0.07 to 0.5 mm under photopic conditions and by 0.44 to 0.7 mm under mesopic conditions, depending on the study.
- Surgically, a large pupil is preferable, but maneuvers to enlarge it may damage the sphincter.
- Concentric EDOF designs are highly pupil-sensitive (a myopic shift with a small pupil); diffractive lenses are less pupil-dependent, but only within a limited range. See the study results.
- A small pupil as a gift: a monofocal lens behind a very small natural pupil behaves optically much like the IC-8, and in an irregular cornea, too, a small pupil reduces HOAs.
- “Fluctuating” vision is not always dry eye: in an aberrated eye, changes in pupil size with lighting cause fluctuations in quality of vision.
Step 3: Lens design
Refractive versus diffractive. In a diffractive lens, the step height determines how energy is split between the foci, and the zone width determines the add power. The measured profiles differ greatly between models, in step height, in step shape and in how the steps change from center to periphery (apodization), and so the light distribution at different pupil sizes differs as well.

Complementary lenses. Pairs from the same family, one for each eye, that differ in how light is distributed between the foci. The measured profiles show steps in similar positions but of different heights and shapes. The clinical evidence so far is mixed, and no advantage has been shown over bilateral implantation of the same lens.

New designs. The spiral lens RayOne Galaxy, whose mechanism the manufacturer has not disclosed in detail, was analyzed independently in the laboratory (Gatinel, Entin & Stern, Biomedical Optics Express 2026). See the optical characterization of the RayOne Galaxy.
Step 4: Simulation, especially in complex eyes
In keratoconus, after corneal refractive surgery and in high ametropia, the outcomes of premium lenses are less predictable, so they are used less often and there is no consensus on their use. Ray-tracing models make it possible to compare several models on the patient's actual cornea, at several pupil sizes, before surgery.


The aim is not to replace clinical judgment, but to make the choice of model a measurement-based process, just as the choice of power has become.
Disclosure: Dr. Stern is a co-founder, CEO and shareholder of AIOLsci.
Checklist before choosing a multifocal or EDOF IOL
- Motivation and expectations; cognitive status and expected ability to adapt.
- Eye health: ocular surface, macula (OCT), optic nerve.
- Topography or tomography: regularity, Q and SA; history of refractive surgery.
- Photopic and mesopic pupil size, before dilation.
- Matching the lens design to the data: pupil sensitivity, interaction with corneal SA, light distribution.
- In a complex eye: simulation, or a monofocal lens that takes advantage of the optics of the cornea and the pupil.
Frequently asked questions
Is an EDOF lens the safe choice after LASIK?
Not automatically. After myopic LASIK, the positive SA of the cornea may neutralize an EDOF lens with negative SA, and after hyperopic LASIK, the negative SA of the cornea may add to that of the lens. The choice should be based on the measured corneal SA and on pupil size.
When should a monofocal lens be considered in a premium candidate?
When the cornea or the pupil already provides depth of focus, for example moderate corneal coma or a very small pupil, and when the eye is complex and the outcome of a multifocal lens is unpredictable. In such situations a monofocal or toric lens may give a good functional result without an additional optical cost.
References
- Stern B, Gatinel D. Presbyopia Correction in Lens Replacement Surgery: A Review. Clin Exp Ophthalmol. 2025;53(6):668-681. Link
- Stern B, Gatinel D. Impact of pupil size and corneal spherical aberrations on the performance of monofocal intraocular lenses: an experimental model. Biomed Opt Express. 2024;15(11):6433-6448. Link
- Stern B, Gatinel D. Comment on: Impact of spherical aberration on visual quality and depth of focus. J Cataract Refract Surg. 2024;50(12):1297-1298. Link
- Gatinel D, Entin A, Stern B. In vitro optical characterization of the RayOne Galaxy spiral extended depth-of-focus intraocular lens using high-resolution Mach–Zehnder interferometry. Biomed Opt Express. 2026;17(7):3984-4001. Link
- Wang L, Dai E, Koch DD, Nathoo A. Optical aberrations of the human anterior cornea. J Cataract Refract Surg. 2003;29(8):1514-1521. Link
- Bottos KM, Leite MT, Aventura-Isidro M, et al. Corneal asphericity and spherical aberration after refractive surgery. J Cataract Refract Surg. 2011;37(6):1109-1115. Link
- Harley SK, Sliney DH. Pupil size in outdoor environments. Health Phys. 2018;115(3):354-359. Link
- Ordiñaga-Monreal E, Castanera-Gratacós D, Castanera F, Fambuena-Muedra I, Vega F, Millán MS. Pupil size differences between female and male patients after cataract surgery. J Optom. 2022;15(2):179-185. Link