Dr. Benjamin SternOphthalmology
For physicians and eye care professionals For physicians · Research

Pupil Size and Corneal Spherical Aberration: How They Change the Performance of Premium IOLs

A summary of an in vitro research program by Dr. Benjamin Stern and Prof. Damien Gatinel, carried out at the Rothschild Foundation Hospital in Paris: the method, the main results by lens, and the conclusions for IOL selection in clinical practice.

In brief

  • The performance of a premium IOL is the product of an interaction between the lens design, the spherical aberration (SA) of the cornea and pupil size – there is no “good lens” in isolation from the eye.
  • Concentric refractive EDOF lenses are very sensitive to the pupil: at a pupil of 2–2.5 mm a myopic shift (“day myopia”) was observed in vitro, more pronounced when the corneal SA is low or negative.
  • The “pupil independence” of a diffractive lens holds only within a defined range of pupil size and corneal SA.
  • High corneal SA (as after myopic LASIK) can neutralize the negative SA of an EDOF lens – and even turn it into a “therapeutic” choice.

Background

IOL selection usually relies on keratometry and axial length, but keratometry samples only the center of the cornea (IOLMaster 500: 2.3 mm; Lenstar: 1.65 and 2.3 mm; IOLMaster 700: 2.5 mm), whereas image quality also depends on the peripheral rays – that is, on the asphericity (Q) and the SA of the cornea. Q and SA are not the same thing:

Asphericity (Q)Corneal SA (µm)
−0.540
−0.26+0.19
0+0.38
+0.26+0.59
+0.54+0.81

For K = 44 D. An average cornea (Q ≈ −0.26) still leaves positive SA; only Q ≈ −0.54 cancels it – a situation that is rare in a natural cornea and more common after hyperopic LASIK.

In the normal population the mean corneal SA is +0.28 µm (6 mm pupil; range 0.055–0.544 µm; Wang et al. 2003). After laser refractive surgery the range is much wider and depends on the amount of correction: after myopic ablation the mean is +0.36 ± 0.17 µm (up to +1.02 µm), and after hyperopic ablation 0.00 ± 0.29 µm (from −0.58 to +0.63 µm; Bottos et al. 2011).

The pupil varies too: in candidates for cataract surgery the photopic pupil ranged from 1.63 to 3.77 mm and the mesopic pupil from 2.43 to 6.82 mm (measured before surgery; Ordiñaga-Monreal et al. 2022). Outdoors the mean pupil is 2.39 mm (1.44–3.03; Harley & Sliney 2018); senile miosis reduces it by about 1.72 mm between ages 20 and 60 in dim light (9 cd/m²), and by less in bright light (Lazar et al. 2024); and after surgery it becomes smaller still – by 0.07 to 0.5 mm under photopic conditions and by 0.44 to 0.7 mm under mesopic conditions, depending on the study (Ordiñaga-Monreal et al. 2022; Kanellopoulos & Asimellis 2014; Pazo et al. 2017).

Methods

Two notes for reading the results. The aperture in the model is the physical diameter of the pupil; the clinically measured pupil (entrance pupil) is about 13% larger, so an aperture of 2 mm corresponds to a measured pupil of about 2.3 mm, and 3 mm to about 3.4 mm. The model is on-axis and monochromatic, and does not include the Stiles-Crawford effect; the values are therefore intended for in vitro comparison and not for use as a nomogram. The method has been published (Stern & Gatinel, Biomedical Optics Express 2024); the results for the premium lenses below were presented at conferences.

Main results

LuxSmart (Bausch & Lomb): myopic shift at a small pupil

In a cornea with no SA, the MTF peak moves to a higher vergence at a pupil of 2–2.5 mm – from about 21 D to about 23 D in the lens plane (in the spectacle plane the difference is roughly 0.7 of that) – meaning that the lens is “stronger” when the pupil is narrow. The HOA map explains this: a central zone of about 2 mm (a combination of 4th-order and 6th-order SA), which at a 2 mm pupil is the entire active lens. The clinical significance: the possibility of a myopic surprise in patients with a small pupil and a cornea with low SA – a finding that, according to the investigators, is also seen in practice.

Through-focus MTF curves of LuxSmart at apertures of 2, 3, 4 and 5 mm
LuxSmart 20 D, ISO-2 corneal model with no SA: at a 2 mm pupil the peak moves to a higher vergence (myopic shift).

AT LISA tri 839M (Zeiss): pupil independence – within a limited range

The lens consists of a central trifocal zone up to 4.3 mm and a peripheral bifocal zone, and the reconstruction from the wavefront clearly showed both zones, consistent with the manufacturer's description. The MTFmax map pointed to a cornea of +0.14 µm as the optimal compensator for the SA of the lens. With this cornea, performance was almost identical between 3 and 4.5 mm – the range the manufacturer emphasizes – but changed significantly at 2–2.5 and at 5–5.5 mm. In a normal cornea with high SA (+0.49 µm) the pupil independence disappeared even between 3 and 4.5 mm. According to studies that link mean MTF to visual acuity, a mean MTF of 0.4 corresponds roughly to 20/20 (0 logMAR) and 0.2 to about 20/32 (0.2 logMAR), so the differences are clinically relevant.

Through-focus MTF curves of AT LISA tri in two corneas and at several apertures
AT LISA tri 839M. Left panel: a cornea that compensates for the SA of the lens (+0.14 µm), apertures of 2–5.5 mm. Right panel: a cornea with high SA (+0.49 µm), apertures of 3–4.5 mm.

AcrySof IQ Vivity (Alcon): “day myopia” and dependence on corneal SA

In an average cornea (+0.28 µm) the vergence of the lens is higher at a pupil of 2–2.5 mm, so a patient with a small photopic pupil can be expected to have “day myopia.” In a cornea with zero SA the shift is more pronounced. In a cornea with high SA (+0.56 µm) the shift is smaller, but the MTF peaks and the EDOF effect in daylight decrease and all the curves move in the myopic direction – an adjustment of the refractive target may be needed. At a 3 mm pupil, depth of focus was greatest in the cornea with low SA (ESCRS 2024 poster, with Maria Rizk and Alain Saad).

Through-focus MTF curves of Vivity in three corneas with different SA
Vivity 20 D in three corneas – SA of 0, +0.28 and +0.56 µm – at apertures of 2–4.5 mm.

PureSee, Vivity and LuxSmart: three routes to refractive EDOF

All three lenses have a central element that creates the EDOF: in PureSee a smooth transition; in Vivity an abrupt transition (described as wavefront shaping, though it may also have a diffractive effect); and in LuxSmart a central bump resembling negative SA, sharper because of 6th-order SA. Vivity and PureSee showed lower MTF peaks but greater depth of focus at a small pupil; LuxSmart behaved as a monofocal at 2 mm, with a power shift that worsens the more negative the corneal SA. At a large pupil the 6 mm optical zone of Vivity (versus 5.5 mm) is an advantage, and LuxSmart achieved better depth of focus in normal eyes and better MTF in corneas with negative SA (ARVO 2024, ESCRS 2025).

The reconstructed surface profiles of PureSee, Vivity and LuxSmart
Reconstruction of the lens profile from the residual wavefront, after removal of the low-order aberrations.

Isopure (BVI) after myopic LASIK: EDOF as a treatment?

At a 4.5 mm pupil, in a cornea after myopic LASIK (SA of +0.91 µm), the MTF peak of Isopure – a lens with high negative asphericity – was much higher and narrower than in an average cornea, and the lens behaved almost as a monofocal. The implication: Isopure may be a therapeutic option for patients after myopic LASIK who suffer from halos – a hypothesis that requires clinical confirmation.

Through-focus MTF curves of Isopure in a normal cornea and in a cornea after myopic LASIK
Isopure, 4.5 mm aperture: a normal cornea (SA +0.28 µm) versus a cornea after myopic LASIK (SA +0.91 µm).

Monofocal lenses too

In aspheric monofocal lenses, too, this interaction is significant (Stern & Gatinel, Biomedical Optics Express 2024).

Clinical significance

  1. Measure the photopic and mesopic pupil before dilation – especially before a concentric refractive EDOF lens.
  2. Perform topography or tomography for every premium lens candidate, and consider the corneal SA, not only the cylinder.
  3. With a small photopic pupil (an aperture of 2.5 mm or less in the model, that is, a measured pupil of up to about 2.8 mm) and a refractive EDOF lens – a myopic shift in daylight is possible; take this into account when choosing the model and the refractive target.
  4. “Pupil independence” is a claim with limits: check over what range of pupil size and SA it was tested.
  5. After LASIK: the SA of the cornea is part of the optical design – it can cancel the EDOF of the lens or amplify it.
  6. A complaint of “fluctuating” vision is not always dry eye: in an aberrated eye, pupil changes alone alter the quality of vision.

The next step: a personalized model for each eye

The research continues at the Biomedical Optics Laboratory that Dr. Stern founded at Hadassah Ein Kerem: a ray tracing model (Zemax OpticStudio) that combines corneal tomography, lens position and tilt from CASIA-2, and IOL profiles measured by optical profilometry (WLI) at nanometer resolution – including asymmetric lenses. According to the plan, the model will be tested against in vivo results from 100 eyes and validated on a further 50.

Separately from his work at Hadassah, Dr. Stern is a co-founder of AIOLsci, an independent company that he founded with Prof. Gatinel and that works on simulating IOL performance in the individual eye.

Disclosure: Dr. Stern is a co-founder, CEO and shareholder of AIOLsci.

Frequently asked questions

Do the in vitro results predict the actual refraction?

They show the direction and the order of magnitude of optical effects, such as a myopic shift at a small pupil, but they do not include neural processing, the Stiles-Crawford effect, chromatic aberrations, tilt and decentration of the lens in the eye, or variability in the effective lens position. They are therefore a tool for understanding and for generating hypotheses, not a nomogram or a substitute for clinical follow-up.

Are LuxSmart or Vivity "no good" at a small pupil?

No. The findings show that performance depends on the anatomy: with a small pupil and a cornea with low SA, a myopic shift is expected that can be taken into account in planning, and in other eyes the same lens may be a very good fit. No lens is good or bad in every eye.

References

  1. Stern B, Gatinel D. Impact of pupil size and corneal spherical aberrations on the performance of monofocal intraocular lenses: an experimental model. Biomed Opt Express. 2024;15(11):6433-6448. Link
  2. Stern B, Gatinel D. Presbyopia Correction in Lens Replacement Surgery: A Review. Clin Exp Ophthalmol. 2025;53(6):668-681. Link
  3. Villemagne N, Joannes L, Stern B, Gatinel D. On bench evaluation of intraocular lenses: performance of a commercial interferometer. Biomed Opt Express. 2024;15(11):6588-6605. Link
  4. Wang L, Dai E, Koch DD, Nathoo A. Optical aberrations of the human anterior cornea. J Cataract Refract Surg. 2003;29(8):1514-1521. Link
  5. Bottos KM, Leite MT, Aventura-Isidro M, et al. Corneal asphericity and spherical aberration after refractive surgery. J Cataract Refract Surg. 2011;37(6):1109-1115. Link
  6. Ordiñaga-Monreal E, Castanera-Gratacós D, Castanera F, Fambuena-Muedra I, Vega F, Millán MS. Pupil size differences between female and male patients after cataract surgery. J Optom. 2022;15(2):179-185. Link
  7. Harley SK, Sliney DH. Pupil size in outdoor environments. Health Phys. 2018;115(3):354-359. Link
  8. Lazar R, Degen J, Fiechter AS, Monticelli A, Spitschan M. Regulation of pupil size in natural vision across the human lifespan. R Soc Open Sci. 2024;11(6):191613. Link

Patient referrals and professional consultation

Physicians and optometrists are welcome to refer patients for a consultation on cataract, premium lenses and refractive surgery, or to contact Dr. Stern with a professional question.

Patient appointments at Hadassah Sharap: 02-677-8899 Direct contact details

The information on this page is intended for professionals and does not replace clinical judgment.