In brief
- In keratoconus, almost all standard formulas lead to a hyperopic surprise, and it grows as the cornea gets steeper. In advanced disease the mean error reaches 2 to 3 D.
- The dedicated formulas, Kane keratoconus and Barrett True-K for keratoconus (preferably with a measured posterior cornea), eliminate most of the bias. Among the older formulas, SRK/T is the best.
- Even with the best formula the spread is wide: in advanced keratoconus only about a quarter of eyes end up within ±0.50 D. A mildly myopic target is preferable to the risk of hyperopia.
- Before biometry: stable disease, rigid contact lenses discontinued, and repeated measurements that agree with one another.
- A toric lens is suitable for mild to moderate, stable keratoconus with a regular, central astigmatic component, in a patient who will not need a rigid contact lens.
Why is keratometry unreliable in keratoconus?
Kane and colleagues (2020) list three reasons, similar to those that apply after refractive surgery but in the opposite direction:
- An abnormal anterior-to-posterior surface ratio. The keratometric index assumes a fixed ratio between the two surfaces. In keratoconus this ratio changes, and the measured K usually overestimates corneal power. The result: an underpowered lens and a hyperopic surprise.
- Prediction of lens position. Formulas that predict the ELP from K “see” a steep cornea and assume an anterior segment different from the one that actually exists.
- The corneal model itself. The keratometer assumes uniform curvature along each meridian and perpendicular principal meridians. In a cone, which is usually inferior rather than central, neither assumption holds.
Added to this is the poor repeatability of the measurements themselves. The repeatability of most devices deteriorates when the maximum K rises above about 55 D, and rigid contact lenses distort the cornea before the measurement.

Before biometry
- Stability. Keratoconus progression slows with age, and by cataract age the disease is usually stable, but it is worth confirming this with repeated tomography. According to a 2018 review, when the disease is not stable, cross-linking (CXL) or intracorneal ring segments before biometry may yield more stable measurements. These treatments are performed in centers that specialize in the cornea.
- Rigid contact lenses. After a rigid lens is discontinued, most of the change in keratometry and corneal thickness occurs in the first week, and the values are stable thereafter (Kumar 2020, 29 eyes). In advanced disease the change is larger. Measure after discontinuation, and confirm with a repeat measurement. As a practical guideline, rigid lenses are stopped at least three to four weeks before the measurements, and the measurement is repeated until it is stable.
- More than one device. Biometer, tomographer and topographer. Treat the discrepancy between them as information about the reliability of the measurement, not as noise from which one value has to be picked.
- A contact lens after surgery. If the patient will need a rigid or scleral lens for good vision, the rigid lens will correct most of the corneal astigmatism. In that situation a toric lens may become a problem rather than a solution.
Which formula?
The largest published series (Kane 2020, 147 eyes) divided the eyes into three stages according to the steep K on the biometer: up to 48 D, 48–53 D, and above 53 D.
| Formula | Stage 1: within ±0.50 D | Stage 2 | Stage 3 | Stage 3: mean error |
|---|---|---|---|---|
| Kane keratoconus | 60.7% | 43.2% | 24.0% | +0.02 D |
| Barrett Universal II | 53.6% | 37.8% | 4.0% | +1.72 D |
| SRK/T | 52.4% | 29.7% | 12.0% | +1.86 D |
| Haigis | 51.2% | 16.2% | 4.0% | +2.43 D |
| Hoffer Q | 52.4% | 16.2% | 0% | +3.02 D |
In non-keratoconic eyes, 75% to 83% of eyes fall within this range. In stage 1 the standard Kane formula is identical to the keratoconus version, because the correction only begins above 48 D.

Three further studies complete the picture:
- Barrett True-K for keratoconus. In a multicenter series from Jerusalem and Maastricht (Vandevenne 2022, 57 eyes), its median absolute error, with a predicted or measured posterior cornea, was 0.10–0.14 D, compared with about 0.5 D for Barrett Universal II and the standard Kane. Its accuracy was similar to that of Kane keratoconus.
- Network meta-analysis (Reitblat 2025, 9 studies, 623 eyes): Barrett True-K with a measured posterior cornea ranked first on all metrics. In severe keratoconus, Kane keratoconus ranked first. A 2026 meta-analysis (7 studies, 530 eyes) reached a similar conclusion, and its authors caution that the data are sparse and heterogeneous.
- In mild keratoconus the differences are small: in the Baylor series (Kozhaya 2023), in eyes with K below 50 D, 58% to 74% of eyes were within ±0.50 D, with no significant difference between formulas. Above 50 D all formulas failed (0% to 18%).
According to the ESCRS guidelines (2024): avoid traditional formulas other than SRK/T, and prefer the keratoconus formulas of Barrett True-K and Kane, especially in advanced stages. Why SRK/T in particular? In a normal eye it tends toward a myopic error when K is high, and this tendency partly offsets the hyperopia of keratoconus. It is an accidental correction, not an intended one.
Choosing the target
The rule: a mild myopic surprise is well tolerated, and a hyperopic one is not. The ESCRS recommends a mildly myopic target. Kane and colleagues suggested, when a third-generation formula is used, a standard target in stage 1, a target of −0.75 to −1.50 D in stage 2, and −2.0 to −3.0 D in stage 3. With a dedicated keratoconus formula the bias is much smaller, so the adjustment required is more modest. But because of the wide spread, a mildly myopic target remains the safe choice.
In advanced keratoconus, when the maximum K is above about 55 D, some authors have suggested calculating with a standard average K value instead of the measured value, while still aiming for a mildly myopic target (Moshirfar 2018, based on a series by Watson). This is an older approach, and the dedicated formulas have largely replaced it.
Toric lenses in keratoconus
When to use one. Mild to moderate, stable keratoconus in which topography shows a regular and relatively central astigmatic component, and the patient is expected to see well with glasses. In the series by Ton and colleagues (2021, 32 eyes), astigmatism fell on average from 2.95 to 0.95 D, and visual acuity improved in all eyes. Ling 2023 also found a significant reduction in astigmatism with a toric lens, with no loss of spherical accuracy.
How to calculate. Here the statistical prediction of the posterior cornea is not valid, so measurement is preferable. In a series of 36 eyes with ectasia (Reitblat 2025), combining several measurements in the Barrett calculator, with a measured posterior cornea, gave the best result, and a predicted posterior cornea was the weakest. The general principles: Toric IOL Calculation.
When not to.
- Predominantly irregular astigmatism, or a cone that is markedly decentered.
- Advanced or unstable disease, or a reasonable likelihood of a future corneal transplant.
- A patient who will need a rigid or scleral contact lens after surgery. According to Moshirfar and colleagues, such a patient should not receive a toric lens.
For similar reasons, multifocal lenses are generally not suitable for an irregular cornea. See Corneal Aberrations and Premium IOLs.
Key points
- Rigid lenses discontinued, repeat tomography, and three measurements that agree with one another before any calculation.
- Use a dedicated formula: Barrett True-K for keratoconus (with a measured posterior cornea) or Kane keratoconus. If neither is available, SRK/T and a myopic target.
- Explain to the patient that accuracy is lower than in a normal eye, especially with an advanced cone, and that a further correction may be needed.
- A toric lens only for a mild to moderate, stable and regular cone, in a patient who will not go back to a rigid lens.
- For patients: Keratoconus.
Frequently asked questions
Which formula is the most accurate in keratoconus?
The dedicated formulas: Barrett True-K for keratoconus, preferably with a measured posterior cornea, and Kane keratoconus, which stood out particularly in advanced keratoconus. If they are not available, SRK/T is the best of the older formulas. In mild keratoconus the differences between formulas are small.
How long should rigid contact lenses be discontinued before biometry in keratoconus?
In a study of 29 eyes, most of the change in keratometry occurred in the first week after discontinuation, and the values were stable thereafter. In advanced disease the change is larger. It is advisable to measure after discontinuation and confirm with a repeat measurement. As a practical guideline, rigid lenses are stopped at least three to four weeks before the measurements, and the measurement is repeated until it is stable.
Can a toric lens be implanted in keratoconus?
Yes, in mild to moderate, stable keratoconus with regular and relatively central astigmatism, in a patient who will not need a rigid contact lens after surgery. With an advanced, irregular or unstable cone, it is better to forgo it.
References
- Kane JX, Connell B, Yip H, et al. Accuracy of intraocular lens power formulas modified for patients with keratoconus. Ophthalmology. 2020;127(8):1037-1042. Link
- Reitblat O, Sella R, Zlatkin R, Bahar I, Lerman TT. Intraocular lens power calculation accuracy in patients with keratoconus: network meta-analysis and systematic review. Clin Exp Ophthalmol. 2025;53(4):356-373. Link
- Vandevenne MM, Webers VS, Segers MH, et al. Accuracy of intraocular lens calculations in eyes with keratoconus. J Cataract Refract Surg. 2023;49(3):229-233. Link
- Kozhaya K, Chen AJ, Joshi M, et al. Comparison of keratoconus specific to standard IOL formulas in patients with keratoconus undergoing cataract surgery. J Refract Surg. 2023;39(4):242-248. Link
- Ton Y, Barrett GD, Kleinmann G, Levy A, Assia EI. Toric intraocular lens power calculation in cataract patients with keratoconus. J Cataract Refract Surg. 2021;47(11):1389-1397. Link
- Moshirfar M, Walker BD, Birdsong O. Cataract surgery in eyes with keratoconus: a review of the current literature. Curr Opin Ophthalmol. 2018;29(1):75-80. Link
- Reitblat O, Zlatkin R, Golan N, et al. Toric intraocular lens calculations in eyes with corneal ectasia. J Refract Surg. 2025;41(6):e558-e568. Link
- Wanten JC, Till V, Findl O, Cochener-Lamard B, Kohnen T, Nuijts RM. ESCRS guideline for cataract surgery 2024: executive summary. J Cataract Refract Surg. 2026;52(9):e1-e13. Link