In brief
- After myopic LASIK or PRK, a standard formula with standard K leads to a hyperopic surprise. After hyperopic treatment the opposite occurs: a myopic surprise.
- Three reasons: the keratometric index is no longer valid, the keratometer measures outside the central zone, and formulas that predict lens position from K get that wrong too.
- Even without historical data there are good methods: Barrett True-K No History, the OCT formula, Haigis-L, Shammas and methods that measure the posterior cornea. The ASCRS calculator runs them side by side.
- The realistic expectation: about 60% to 70% of eyes within ±0.50 D when history is available, and slightly fewer without it. This is what the patient should be told before surgery.
- After RK: an unstable cornea, a tendency toward hyperopia after surgery, and a mildly myopic target. Do not rush into an IOL exchange.
Why do the standard formulas go wrong?
Axial length remains reliable after refractive surgery. The problem is corneal power, and through it lens position as well. Savini and Hoffer summarize three sources of error:
- Keratometric index error. The keratometer measures the radius of the anterior surface and converts it to diopters with a fixed index (usually 1.3375), which assumes a fixed ratio between the anterior and posterior surfaces. Myopic ablation flattens only the anterior surface, the ratio changes, and the measured K is higher than the true corneal power. After hyperopic ablation it is lower.
- Radius error. The keratometer measures on a paracentral ring. With a small or decentered optical zone it may measure outside the treated zone. With an optical zone of 6 mm or more, this error is negligible.
- Formula error (ELP). Third-generation formulas (SRK/T, Holladay 1, Hoffer Q) predict lens position from K. A cornea that is flat after myopic treatment “looks” like an eye with a shallow chamber, and the predicted ELP is too anterior. This is why the Double-K method (Aramberri) was developed: the pre-laser K to predict position, and the current K to calculate power. Formulas that predict position without K, such as Haigis, are exempt from this error.
After myopic treatment all three act in the same direction: the calculated IOL power is too low, and the result is hyperopic. The larger the laser correction, the larger the error.

Before the calculation: know what was done
- Which surgery? Myopic ablation flattens the central cornea, and hyperopic ablation steepens it. With RK, topography shows flattening and irregular astigmatism. A LASIK flap and RK incisions are visible at the slit lamp, but PRK can usually be identified only on topography. There have been cases in which the patient remembered “laser for myopia” and the cornea showed hyperopic treatment.
- Centration of the ablation. A decentered ablation creates irregular astigmatism, which limits corrected vision and the choice of lens.
- Historical data. If K and refraction from before the laser are available, along with a stable refraction after it (before the cataract shifted it), they are worth using. The AAO and the ISRS developed a “K card” (K-Card) on which the patient keeps these data.
- The ocular surface. A cornea that has had laser surgery is often dry. Measuring on an unstable surface adds error to a problem that already exists (see Pitfalls in Biometry).
The methods: with and without history
| Method | What is needed | The principle |
|---|---|---|
| Clinical history method | Pre-laser K and the change in refraction | The correction is subtracted from the original K. Highly vulnerable to erroneous data, and according to the BCSC and Savini-Hoffer no longer considered the gold standard |
| Masket, Adjusted Atlas and similar methods | The change in refraction from the laser | A regression-based adjustment of the IOL power or of K according to the amount of correction |
| Barrett True-K | With or without history (No History) | A version of Barrett Universal II with an adjusted K. Unpublished. Also built into some biometers, and the TK version makes use of the posterior corneal measurement |
| Haigis-L | No history | Haigis (lens position from ACD) with a correction curve for the measured K. Separate versions for myopic and hyperopic treatment |
| Shammas-PL (and PHL after hyperopic treatment) | No history | A linear correction of K, in a formula that does not predict position from K |
| The OCT formula | Corneal OCT | Net corneal power from both surfaces, with lens position based on ACD, lens thickness and axial length |
| Total keratometry, tomography, ray tracing | Measurement of the posterior surface | Measuring instead of assuming. Requires adapted constants and a formula that does not predict position from K. With the newer formulas (EVO 2.0, Barrett True-K TK) the data are promising |
| Intraoperative aberrometry | Measurement of the aphakic eye in the operating room | According to the 2024 AAO review, accuracy similar to Barrett True-K and slightly better than other methods |
The ASCRS calculator (developed by Hill, Wang and Koch) takes the biometer, topography and historical data, and displays the results of the different methods side by side, together with the mean and the range. According to the BCSC it is probably the best option for preoperative calculation, and the ESCRS guidelines (2024) also recommend the methods it contains. A wide spread between the methods is a warning sign in itself: check the measurements before choosing a number.
What accuracy can be expected?
After myopic LASIK or PRK, in most series 60% to 70% of eyes fall within ±0.50 D when historical data are available, and 50% to 60% without them, up to about 70% with ray tracing or aberrometry (Savini & Hoffer 2018). In an eye that has not had surgery, in the large series, the rate is about 70% to 83% with a modern formula (see IOL Power Calculation Formulas).

Two network meta-analyses (Wei 2021, 19 studies, 1,217 eyes; Pan 2024, 24 studies, 1,172 eyes) found that Barrett True-K, the OCT formula and Masket (and, in the first, intraoperative aberrometry as well) are more accurate than Haigis-L within ±0.50 D. In Pan 2024, Wang-Koch-Maloney was the weakest, and after RK no difference was found between the methods. Two practical conclusions: no single method always wins, and the differences between the leading methods are smaller than the spread of each one of them.
After hyperopic treatment
Here the error is reversed: the measured K is lower than the true value, and the danger is a myopic surprise. The data are far more limited. In the published series (21 to 107 eyes), no method was found to be clearly superior among Barrett True-K, Haigis-L, Masket and the newer methods, although in the largest series (Boccia 2025) Shammas was less accurate. Hyperopic ablation makes the cornea more prolate (negative spherical aberration), so in these eyes a lens without negative spherical aberration is considered, and caution is required when choosing a multifocal lens (Corneal Aberrations and Premium IOLs).
After RK
- The cornea itself is unstable. In the PERK study, a hyperopic shift of one diopter or more was measured in 43% of eyes between six months and ten years after RK, and in some patients vision fluctuates during the day.
- Two errors in different directions. After RK the posterior surface flattens more than the anterior surface, and in addition the optical zone is small (about 3 mm). It is therefore preferable to use central power values from topography, in a Double-K formula, and not SimK.
- Historical data help here. In the series by Turnbull and colleagues (2020, 52 eyes), True-K with history brought about 75% of eyes within ±0.50 D, and without history, True-K No History and Haigis reached 69%.
- A myopic target. The incisions swell after cataract surgery, the cornea flattens and a temporary hyperopia results, lasting weeks to months. When standard formulas are used, the BCSC recommends a target of −0.50 to −1.50 D. Modern formulas try to correct for this tendency, but even with them a slight myopic residual is preferable to hyperopia.
- Do not rush into an exchange. With a refractive surprise after RK, wait for the cornea and the refraction to stabilize, sometimes for months, before an IOL exchange. After LASIK or PRK the cornea is more stable, and a correction can be planned earlier (see Management of Refractive Surprise).
After SMILE the principles are similar, but the data are very limited and based mainly on comparison with ray tracing. Here too it is advisable to use several methods in parallel.
Key points
- Topography or tomography before any calculation after refractive surgery: identify the type of treatment, the centration and the irregular astigmatism.
- Calculate with several methods (the ASCRS calculator) and look at the spread, not only at the mean. If Barrett True-K is available on the biometer, it is a good starting point and avoids transcription errors.
- After myopic treatment, when the methods disagree, it is better to be on the myopic side of the range.
- Explain to the patient in advance that the accuracy is lower than what they knew from their laser surgery, and that a further correction may be needed. Page for patients: Cataract Surgery After Laser Eye Surgery.
Frequently asked questions
The patient has no data from before the laser. Can the calculation be reasonably accurate?
Yes. Methods that do not need history, such as Barrett True-K No History, the OCT formula and Haigis-L, bring about 50% to 65% of eyes within half a diopter after myopic laser surgery, slightly fewer than with history. It is advisable to compare several methods in the ASCRS calculator.
Which target should be chosen after RK?
A mildly myopic target. After cataract surgery the incisions swell and the cornea flattens, which creates a temporary hyperopia. When standard formulas are used, the BCSC recommends a target between −0.50 and −1.50 D.
Is intraoperative aberrometry preferable after refractive surgery?
According to the 2024 AAO review, its accuracy is similar to that of Barrett True-K and slightly better than that of older methods. It can be used for confirmation, but it does not replace a careful preoperative calculation.
References
- Savini G, Hoffer KJ. Intraocular lens power calculation in eyes with previous corneal refractive surgery. Eye Vis (Lond). 2018;5:18. Link
- American Academy of Ophthalmology. Basic and Clinical Science Course, Section 13: Refractive Surgery, chapter 8 (Considerations after refractive surgery); and Section 11: Lens and Cataract, chapter 7. San Francisco: AAO; 2022-2023.
- Abulafia A, Hill WE, Koch DD, Wang L, Barrett GD. Accuracy of the Barrett True-K formula for intraocular lens power prediction after laser in situ keratomileusis or photorefractive keratectomy for myopia. J Cataract Refract Surg. 2016;42(3):363-369. Link
- Pan X, Wang Y, Li Z, Ye Z. Intraocular lens power calculation in eyes after myopic laser refractive surgery and radial keratotomy: Bayesian network meta-analysis. Am J Ophthalmol. 2024;262:48-61. Link
- Wei L, Meng J, Qi J, Lu Y, Zhu X. Comparisons of intraocular lens power calculation methods for eyes with previous myopic laser refractive surgery: Bayesian network meta-analysis. J Cataract Refract Surg. 2021;47(8):1011-1018. Link
- Turnbull AMJ, Crawford GJ, Barrett GD. Methods for intraocular lens power calculation in cataract surgery after radial keratotomy. Ophthalmology. 2020;127(1):45-51. Link
- Boccia R, Lanza M, Luciano G, et al. Evaluation of reliability of formulas for intraocular lens power calculation after hyperopic refractive surgery. J Clin Med. 2025;14(6):1990. Link
- Pantanelli SM, Hatch K, Lin CC, et al. Intraoperative aberrometry versus preoperative biometry for intraocular lens power calculations: a report by the American Academy of Ophthalmology. Ophthalmology. 2025;132(2):238-252. Link