How to Read Optical Bench Data for Intraocular Lenses

MTF graphs appear in every brochure and every paper on premium lenses, but each graph answers a very specific question. What the metrics mean, which details to look for in the fine print, and how to avoid wrong conclusions, with examples from Dr. Stern's work.

In brief

  • Every optical bench graph is an answer to a specific question: which model eye, which aperture, which wavelength, which spatial frequency and which lens power. Without these five details, graphs cannot be compared.
  • MTF at a single frequency (for example MTF50) shows fine detail and separates foci well; MTFa (or mean MTF over 0–50 lp/mm) correlates better with visual acuity.
  • A mean MTF of about 0.4 corresponds roughly to 0.0 logMAR and 0.2 to about 0.2 logMAR (20/32). Beyond the upper threshold, a higher peak does not necessarily mean better acuity.
  • The ISO 1 model eye (an aberration-free cornea) penalizes lenses that correct SA, and the ISO 2 model (a cornea with average SA) suits an average eye only. After LASIK neither is representative.
  • Differences smaller than the variability of the instrument are noise. On the instrument used for the measurements in Dr. Stern's work, the smallest detectable difference in power is about 0.2 D.

MTF: how much contrast gets through, and at what frequency

The MTF describes how much of the contrast of a stripe pattern passes through the optical system, as a function of the density of the stripes (spatial frequency, in lp/mm at the retina). In a model eye with a lens of medium power, 100 lp/mm corresponds to about 30 cycles per degree (roughly 20/20), and 50 lp/mm to about 15 cycles per degree (roughly 20/40). The MTF also depends on the lens power, so lenses are compared at the same power, usually around 20 D.

Graph of MTF as a function of spatial frequency: a diffraction-limited curve and a curve with aberration, with the area under the curve up to 50 lp/mm shaded
MTF as a function of spatial frequency, at a 3 mm aperture and in green light: a diffraction-limited system (gray) and the same system with spherical aberration (blue). MTF50 is a single value on the curve; MTFa is the shaded area up to 50 lp/mm, and mean MTF is the area divided by 50. Calculation prepared for this page.

Through-focus: the metric that really matters

In a through-focus graph the MTF (at a single frequency, or the mean) is computed in different focal planes and displayed as a function of defocus in diopters. It is the optical equivalent of a clinical defocus curve. Three pitfalls on the horizontal axis:

Mean MTF curve as a function of vergence, with MTFmax, the refractive error tolerance range, and DOFmin and DOFmax marked
The metrics that Dr. Stern and Prof. Gatinel defined on the mean MTF curve (Biomed Opt Express 2024): MTFmax (the height of the peak), RET (the range over which the mean MTF is above 0.4, the “landing zone”) and the depth of focus from the 0.4 threshold down to 0.2, measured from each of the two ends of the RET (DOFmax and DOFmin). From his ASCRS 2024 poster.

Other metrics

MetricWhat it measuresWhat to watch for
MTF50 / MTF100Contrast at a single frequencySeparates the foci of a multifocal lens well; sensitive to aberrations; correlates less well with acuity
MTFa or mean MTF (0–50 lp/mm)The area under the MTF curveCorrelates well with visual acuity; the American standard translates it into predicted acuity
Strehl ratioThe peak of the system's PSF divided by the peak of the PSF of the same system when diffraction-limitedA single number; suited mainly to small aberrations; “visual Strehl” weights for the sensitivity of the visual system
PSFThe image of a point of lightShows halo and light scatter around the focus; the American standard for EDOF includes a laboratory halo test
Target image (USAF 1951, letters)An image of a target through the lensIntuitive but qualitative; depends on the exposure, the camera and the choice of frame
Simulated defocus curveMTFa translated into predicted acuityThe formula was derived from studies of particular eyes; for pupil-dependent lenses, an adjustment for pupil size is more accurate

The link between MTF and acuity: according to studies that related MTFa to acuity, including the formula adopted in the annex of ANSI Z80.35 (VA = 0.085 / MTFmean − 0.21), a mean MTF of 0.4 corresponds to about 0.0 logMAR and 0.2 to about 0.2 logMAR. Above 0.4, additional MTF hardly improves acuity. A comparison of “a peak of 0.8 versus 0.7” between two monofocal lenses is therefore probably clinically meaningless, whereas 0.25 versus 0.15 at intermediate can be a difference of about two lines.

The model eye: ISO 1, ISO 2 and what lies in between

Limitations shared by the ISO models: there is no angle alpha, the pupil is centered, the retina is flat, and the measurement is on-axis only. In other words, tilt and decentration are not tested. See IOL Tilt and Decentration.

Aperture and wavelength

An example: the same four lenses, four graphs

Four through-focus graphs for four lenses: MTF at 50 lp/mm and MTFa, at apertures of 3 and 4.5 mm
Four lenses (a monofocal, Isopure, a diffractive trifocal and the spiral Galaxy), in the Arizona pseudophakic eye model and in 543 nm light: MTF50 (left) and MTFa (right), at an aperture of 3 mm (top) and 4.5 mm (bottom). From the paper by Gatinel, Entin and Stern (Biomed Opt Express 2026).

What you see: with MTF50 the trifocal lens shows separate peaks, and the EDOF lens looks weak everywhere; with MTFa the same EDOF lens shows a continuous “plateau” over about 2 D. Going from 3 to 4.5 mm, the distance peak of Galaxy rises and Isopure weakens. There is no single “correct” graph here: each panel answers a different question.

How the data mislead

  1. A model eye chosen to suit the lens. A lens that corrects SA will look excellent with a cornea of average SA and less good in the ISO 1 model, and a spherical lens the reverse. In Dr. Stern's work on 168 model eyes, earlier studies that contradicted one another were explained by the choice of model eye.
  2. A design instead of a lens. A simulation of the design is not a manufactured lens. In the validation work for the interferometer, an aspheric lens with a high conic constant showed a peak 0.06–0.10 MTF units lower than the design, which was attributed to manufacturing deviation.
  3. A difference smaller than the noise. In Dr. Stern's repeatability study, the minimal detectable difference (MDD) was 0.216 D in power and 0.053 in MTF50 at the distance peak. A smaller difference than this between two lenses is not a finding.
  4. Different powers. The MTF depends on the lens power, and the SA of a spherical lens increases with power. A comparison between 18 D and 24 D is not a fair comparison.
  5. A “photopic” aperture of 3 mm. In cataract patients the pupil outdoors is often smaller than that, and with refractive lenses the behavior at a pupil of 2–2.5 mm can be very different.
  6. On-axis only. Aspheric and multifocal lenses are sensitive to tilt and decentration, which do not appear in the graph.
  7. Contrast is not halo. A high MTF does not guarantee few halos; for that you need the PSF or a halo test.

The measurements in Dr. Stern's work were made with a NIMO TEMPO interferometer. Patient version: How Lenses Are Tested in the Lab.

Frequently asked questions

Which metric is closest to the patient's visual acuity?

MTFa, or mean MTF between 0 and 50 lp/mm, has been found in studies to correlate well with visual acuity, and the American standard uses it to predict acuity. It too depends on the model eye and the aperture, and for pupil-dependent lenses an adjustment for pupil size is more accurate.

Can graphs from two manufacturers be compared?

Only if all the conditions are identical: model eye (and corneal SA), aperture, wavelength, spatial frequency, lens power and the same type of bench. In most cases they are not identical, so an independent comparison in the same laboratory is more reliable.

References

  1. Stern B, Gatinel D. Impact of pupil size and corneal spherical aberrations on the performance of monofocal intraocular lenses: an experimental model. Biomed Opt Express. 2024;15(11):6433-6448. Link
  2. Villemagne N, Joannes L, Stern B, Gatinel D. On bench evaluation of intraocular lenses: performance of a commercial interferometer. Biomed Opt Express. 2024;15(11):6588-6605. Link
  3. Stern B, Saad A, Flamant R, Joannes L, Gatinel D. Intra-observer and inter-observer variability of intraocular lens measurements using an interferometry metrology device. Diagnostics. 2024;14(2):216. Link
  4. Gatinel D, Entin A, Stern B. In vitro optical characterization of the RayOne Galaxy spiral extended depth-of-focus intraocular lens using high-resolution Mach–Zehnder interferometry. Biomed Opt Express. 2026;17(7):3984-4001. Link
  5. Vega F, Millán MS. Predicting visual acuity in pseudophakic eyes with pupil-dependent extended depth of focus intraocular lenses. J Refract Surg. 2025;41(8):e831-e838. Link
  6. Monteiro DWL, Trindade BLC, eds. Advances in Ophthalmic Optics Technology. Bristol: IOP Publishing; 2022 (chapters on optical quality metrics and IOL eye models). Link
  7. Faria-Ribeiro M, Villemagne N, Joannes L. Physiological chromatic model eyes for IOL characterization. Biomed Opt Express. 2026;17(5):2512-2522. Link
  8. Moschitta D, Lutkenhaus J, Choi MT, Lee S. Prediction of monocular and binocular visual acuity for the pseudophakic patients using simulated modulation transfer function. Clin Ophthalmol. 2026;20:620493. Link

Patient referrals and professional consultation

Physicians and optometrists are welcome to refer patients for a consultation on cataract, premium lenses and refractive surgery, or to contact Dr. Stern with a professional question.

Patient appointments at Hadassah Sharap: 02-677-8899 Direct contact details

The information on this page is intended for professionals and does not replace clinical judgment.