In brief
- In cataract surgery, an intraocular lens (IOL) replaces the natural lens that has turned cloudy. It stays in the eye for life, and it cannot be “tried on” before surgery the way glasses can.
- In an optical laboratory the lens is tested outside the eye: laser light is passed through it, its “optical fingerprint” is measured, and a computer simulates how it will behave inside an eye.
- The measurements show that the same lens behaves differently in different eyes: the size of the pupil and the shape of the cornea change the sharpness of the image and the range of vision.
- So there is no “best lens” for everyone, only a good match between a lens and an eye. The tests done before surgery are the basis for that match.
What is an intraocular lens, and why is it tested in the lab?
In cataract surgery, the natural lens that has turned cloudy is removed and a clear artificial lens is implanted in its place: an intraocular lens. Today there is a wide variety of lenses: monofocal lenses for distance vision, “monofocal plus” lenses, extended depth of focus (EDOF) lenses, and multifocal and trifocal lenses that also provide near vision. You will find an explanation of the different types on the page Types of Intraocular Lenses.
The problem is that a patient cannot try a lens before choosing it. Dr. Benjamin Stern compares it to buying shoes that you will have to wear for the rest of your life, with no option of exchanging them: the size (the lens power) matters, but it is not enough to tell you whether the shoe will fit. In addition, new models reach the market every year, faster than the time it takes to test them in large clinical studies. Laboratory testing makes it possible to examine a lens independently, rather than relying only on the information the manufacturer publishes.

How is a lens measured in the lab?
Dr. Stern performed the measurements during his fellowship at the Rothschild Foundation Hospital in Paris, together with Prof. Damien Gatinel, on a device called the NIMO TEMPO. It is not a conventional “optical bench” but an interferometer: a device that measures, with great precision, the shape of the light wave that passes through the lens. The process looks like this:
- Preparation: the lens is placed inside a small measurement cell filled with fluid.
- Measurement: a green laser beam passes through the lens, and the device photographs a pattern of rings of light and shadow (interference fringes) that encodes its exact shape.
- Reconstruction: the computer translates the pattern into a “wavefront” (wavefront), a kind of optical fingerprint of the lens.
- Simulation: this fingerprint is “implanted” in a computer-modeled eye, with corneas and pupils of different sizes, and the image that will form on the retina is calculated.

The advantage of the method is its flexibility: after a single measurement, the same lens can be tested in dozens or even hundreds of different computer-modeled eyes. In addition, the measurement captures the lens as it was actually manufactured, including small manufacturing imperfections, and not just its design on paper.
What do we learn from the measurement?
- The lens power in diopters, and whether it has cylinder (as in toric lenses for correcting astigmatism).
- Image quality: the standard measure is called MTF. Put simply, it shows how much of the contrast of fine details the lens transfers to the retina. A higher value means a sharper, clearer image.
- Range of vision: a through-focus curve shows image quality at every distance. A monofocal lens has one peak, usually for distance. A trifocal lens has three peaks: for distance, intermediate (computer) and near (reading). An EDOF lens has a lower, wider peak.
- The structure of the lens surface: in multifocal lenses it is possible to reconstruct the tiny “steps,” just a few microns high, that split light into different focal points.


Same lens, different eyes
Manufacturers usually publish only four graphs for each lens: two types of cornea and two pupil sizes. But not all eyes resemble the average. In his research, Dr. Stern built 168 model eyes, with 21 corneal shapes and eight pupil sizes, and tested every lens in them. It turned out that two factors change the result a great deal.
Pupil size
The pupil is the black opening in the center of the iris, and it works like the aperture of a camera: in bright light it is small, and in the dark it gets larger. Its size varies widely between people, and it gets smaller with age. In a large population study in Germany the median pupil in dim lighting was about 4.2 mm; at age 70 it ranges in most people between about 3 and 5 mm. A small pupil improves depth of field, like a pinhole, but lets in less light. A large pupil lets in more light, but also more optical distortions.
With advanced lenses the effect is especially strong. In Dr. Stern's measurements, a trifocal lens tested with a 2 mm pupil behaved like an EDOF lens, because the small pupil “hid” some of its rings. With some EDOF lenses that have a special central zone, a very small pupil in daylight can cause mild nearsightedness (“daytime nearsightedness”), a finding that, according to Dr. Stern, is sometimes also seen in the clinic.
The shape of the cornea
The cornea, the clear window at the front of the eye, is responsible for about two-thirds of the eye's focusing power. A healthy cornea is steeper in the center and flatter toward the edges, and this structure determines the amount of “spherical aberration”: a distortion in which light rays passing through the periphery of the eye are focused slightly differently from rays passing through the center. The amount varies from person to person, and it changes a great deal after laser vision correction (LASIK or PRK). Different lenses are designed to add to this aberration, cancel it or leave it unchanged, which is why the same lens can give a sharp image in one eye and a less sharp image in another. More information on the page Cataract Surgery After Laser Eye Surgery.
“There are no good or bad lenses in general. There is a good match, or a less good match, between a lens and an eye.” (Dr. Benjamin Stern)
How accurate are the measurements?
Before relying on a device, you have to test it. In a study published in 2024, Dr. Stern and his colleagues tested eight different lenses, with powers from 8 to 34 diopters. Three ophthalmologists measured each lens ten times. The measurement deviation for lens power was less than a tenth of a diopter (0.078 D), and the device reliably distinguished between lenses whose power differed by only about 0.2 diopters. That is less than half the usual interval between two adjacent lens powers (typically half a diopter). In another study, from the same year, the device's results were compared with the optical design of the lenses and with a different type of measuring device, and good agreement was found.
What the lab cannot predict
A laboratory measurement is an important tool, but it does not replace the clinical exam and the surgeon's judgment. It tests the optics only. In a real eye other things matter too: the brain, which learns to adapt to a new lens (neuroadaptation), the condition of the retina, dry eye, the exact position of the lens in the eye after surgery, and each person's needs and habits. So laboratory results are not always identical to what is seen in the clinic, and they do not guarantee any particular outcome.
What does this mean for you as a patient?
- The tests before surgery matter: besides the measurements used to calculate the lens power (biometry), mapping of the cornea (topography or tomography) and attention to pupil size help estimate how a particular lens will behave in your eye. See Before Cataract Surgery.
- Mention previous surgeries: laser vision correction, even if it was done many years ago, changes the shape of the cornea and the considerations in choosing a lens.
- Talk about your needs: reading, computer work, activities after dark. Choosing a lens is a balance between sharpness, range of vision and phenomena such as halos around lights.
- Ask questions: what are the advantages and drawbacks of each lens in your eye? You do not have to decide alone; this is a decision you and your surgeon make together.
The research continues
In 2025 Dr. Stern founded the Biomedical Optics Laboratory at Hadassah Ein Kerem, which studies the interaction between the cornea, the pupil and the intraocular lens using computer modeling.
Separately from his work at Hadassah, Dr. Stern is a co-founder of AIOLsci, an independent company that he founded together with Prof. Damien Gatinel and that began during his fellowship in Paris. The company is developing software for the optical comparison of lenses, and in 2026 it won an award from the International Society of Refractive Surgery.
Frequently asked questions
Is the lens that will be implanted in my eye tested in the lab before surgery?
No. Laboratory studies test lens models, in order to understand how each model behaves under different conditions. Every lens that is implanted is manufactured and tested by the manufacturer according to quality standards, and arrives sterile in sealed packaging.
Can a laboratory test guarantee me a particular level of vision after surgery?
No. The lab tests only the optics of the lens. The actual result also depends on the condition of the retina and cornea, on dry eye, on the position of the lens and on how the brain adapts. The information from the lab helps in understanding the advantages and drawbacks of each type of lens, and the decision is made after a full exam and a conversation with the surgeon.
Why does pupil size matter when choosing a lens?
The pupil determines which part of the lens takes part in forming the image. With multifocal lenses and EDOF lenses, a very small or very large pupil can change the sharpness of the image and the range of vision. That is why the doctor also takes pupil size into account in the tests before surgery.
I had laser vision correction years ago. Does that change anything?
Yes. Laser surgery changes the shape of the cornea, and this affects both the calculation of the lens power and the question of which type of lens will be suitable. It is important to tell your doctor about it and, if possible, to bring information about the previous surgery. More details on the page Cataract Surgery After Laser Eye Surgery.
Who carried out these studies?
Dr. Benjamin Stern, together with Prof. Damien Gatinel, during his fellowship at the Rothschild Foundation Hospital in Paris. The results were presented at international conferences and published in scientific journals, and the research continues at the Biomedical Optics Laboratory he founded at Hadassah.
References
- Stern B, Saad A, Flamant R, Joannes L, Gatinel D. Intra-Observer and Inter-Observer Variability of Intraocular Lens Measurements Using an Interferometry Metrology Device. Diagnostics. 2024;14(2):216. Link
- Stern B, Gatinel D. Impact of pupil size and corneal spherical aberrations on the performance of monofocal intraocular lenses: an experimental model. Biomed Opt Express. 2024;15(11):6433-6448. Link
- Villemagne N, Joannes L, Stern B, Gatinel D. On bench evaluation of intraocular lenses: performance of a commercial interferometer. Biomed Opt Express. 2024;15(11):6588-6605. Link
- Gatinel D, Entin A, Stern B. In vitro optical characterization of the RayOne Galaxy spiral extended depth-of-focus intraocular lens using high-resolution Mach–Zehnder interferometry. Biomed Opt Express. 2026;17(7):3984-4001. Link
- Kiel M, Grabitz SD, Hopf S, et al. Distribution of Pupil Size and Associated Factors: Results from the Population-Based Gutenberg Health Study. J Ophthalmol. 2022;2022:9520512. Link