In brief
- The capsular bag and zonules have three roles: stabilization, accommodation and centration of the lens. Without them the IOL needs another anchor: the angle, the iris or the sclera.
- Risk factors for subluxation and dislocation: severe PXF, high myopia, prior vitrectomy, trauma, Marfan syndrome and homocystinuria, RP, and complicated surgery with posterior capsule rupture.
- The options: ACIOL, iris fixation (sutures or iris-claw), scleral fixation (sutures, devices, or externalization of the haptics with the Yamane technique).
- The AAO review (2020) found no clear superiority of any one technique. The choice is tailored to the eye, the existing IOL and the surgeon.
- Even a well-fixated IOL may be tilted. AS-OCT makes it possible to quantify this, and it matters when choosing a multifocal or aspheric lens.
The capsular bag and zonules: why do they matter so much?
“A good bag is the cataract surgeon's best friend.” The zonules of Zinn are spread around the lens, on its anterior, equatorial and posterior aspects. Each fiber is 5–30 µm in diameter and is built of microfibrils 8–10 nm in diameter, composed mainly of fibrillin 1 and fibrillin 2. Hence the link to Marfan syndrome. The microfibrils are both strong and elastic, and they absorb shocks.
The capsular bag and zonules have three functions: stabilization, accommodation and centration. Their uniform circumferential arrangement allows the lens to change shape without creating astigmatism. Conversely, zonules of unequal length, or zonules that are partly missing or damaged, cause tilting (tilt) and displacement (decentration) of the lens, even when no zonules at all are visible at the slit lamp. When zonules are visible at the slit lamp in a given quadrant, the lens is already clearly off axis, and this may cause refractive changes and high astigmatism.
With AS-OCT, the tilt and decentration of the natural lens can be measured before surgery, and they partly predict the position of the IOL afterward. In a Japanese study (Kimura 2017), in 17 eyes scanned before and after surgery, the correlation was good for tilt (ICC 0.73–0.91) and moderate for decentration (ICC 0.41–0.56). In a larger study with an SS-OCT biometer (Wang 2019, 333 patients), IOL tilt was on average 1.2° greater than that of the natural lens, and significantly correlated with it. See Uses of AS-OCT in Cataract Surgery.
Risk factors for subluxation and dislocation
| Compromised zonular support | Absent or insufficient capsular support |
|---|---|
| PXF, in severe cases | Complicated cataract surgery with posterior capsule rupture |
| High myopia | Trauma |
| Prior vitrectomy | |
| Prior trauma | |
| Marfan syndrome, homocystinuria and others | |
| Retinitis pigmentosa |
Early signs worth looking for before surgery: phacodonesis, iridodonesis, PXF material, an asymmetric anterior chamber, and zonules visible in one quadrant. For an overview written for patients, see Dislocated Intraocular Lens.
Fixation options
| Method | Anchor | Practical notes |
|---|---|---|
| Exchange for an ACIOL | Anterior chamber angle | Technically simple and readily available. Requires a deep chamber, a healthy endothelium and an open angle, and is less suitable in glaucoma or with PAS. |
| Iris suture fixation | Iris | For a three-piece IOL whose haptics are sutured to the iris from behind it. Requires an intact iris. |
| Iris-claw | Iris (enclavation) | Anterior or retropupillary. Does not depend on capsular support. |
| Scleral sutures or devices | Sclera | For example an IOL with suture eyelets (such as the Akreos AO60) with a Gore-Tex suture (off-label use), including IOL exchange. Another variation that was presented: a triple support suture for the lens (“clover leaf”). |
| Sutureless externalization of the haptics (Yamane) | Sclera (intrascleral) | Three-piece IOL. The haptics are externalized through thin needles and fixed with a thermally created bulb (“flange”). Usually combined with vitrectomy. Variations were also presented, such as the snare technique. |
How to choose?
The AAO's 2020 Ophthalmic Technology Assessment report (Intraocular Lens Implantation in the Absence of Zonular Support: An Outcomes and Safety Update) found no clear superiority of any of the accepted techniques. The report reviewed 45 studies, most of them retrospective series, and visual outcomes were similar across techniques. Each technique has its own complication profile: compared with ACIOL, CME was reported more often with an iris-sutured PCIOL and with scleral suturing using polypropylene 8-0, retinal detachment was roughly twice as common with the sutured techniques (other than Gore-Tex) as with the sutureless techniques, and chronic uveitis was reported less often with the non-ACIOL techniques. In the original Yamane series (2017, 100 eyes), mean IOL tilt was 3.4° ± 2.5°, and the common complications were capture of the optic by the pupil (iris capture, 8%) and transient vitreous hemorrhage (5%). In practice, the decision rests on:
- The cornea and the angle: endothelial status, chamber depth, glaucoma and PAS (considerations against an ACIOL).
- The iris: tissue integrity and pupil motility (a consideration in iris fixation).
- The vitreous and retina: whether a vitrectomy is required, whether the IOL has dropped posteriorly.
- The existing IOL: a three-piece lens can sometimes be refixated. A single-piece acrylic lens is not suitable for the sulcus or for iris fixation, and will usually be exchanged or fixated together with the bag.
- The surgeon's experience and the availability of lenses and devices.
Tilt after fixation, and why it matters
Multifocal lenses are built as concentric rings around the center of the IOL, on the assumption that the visual axis passes through it. Tilt or decentration undermines this assumption. A simulation study presented at the meeting (Madrid-Costa 2012, adaptive optics, 10 young eyes) tested two aspheric and two spherical lenses, centered, with decentration of 0.2 and 0.4 mm and with tilt of 2° and 4°. When the lenses were centered there was no difference between them. Decentration of as little as 0.2 mm slightly reduced high-contrast visual acuity, and at low contrast the reduction was mainly with the aspheric lenses, which were also more sensitive to tilt. The differences are small, on the order of one to two lines, and according to the classic work of Holladay an aspheric lens is still preferable to a spherical one as long as decentration is less than 0.4 mm and tilt is less than 7°. Beyond that, a spherical lens, or a neutral aspheric one, is more forgiving. So when significant tilt is expected, it is worth thinking about which monofocal lens to choose, and thinking twice about a multifocal lens. A high chord mu adds to the problem as well: Chord Mu and Angle Alpha.
Case: a scleral-sutured IOL and vertical tilt
A woman in her 70s was referred to the emergency room because of high intraocular pressure in the setting of PXF glaucoma and cataract. At surgery, severe phacodonesis and vitreous in the anterior chamber were found. A pars plana vitrectomy was performed, and a three-piece IOL was sutured to the sclera.
On postoperative AS-OCT, the lens is centered and level in the horizontal section, but the vertical section shows obvious tilt. The lesson: successful secondary fixation does not guarantee a straight lens, and AS-OCT is the simple way to quantify this.


Key points
- Preserve the bag: identify signs of zonular weakness in advance, and plan accordingly (capsular tension ring, devices, an appropriate lens).
- Tilt of the natural lens before surgery partly predicts IOL tilt afterward.
- There is no “best technique”: match the technique to the eye, the existing IOL and your own experience.
- In an eye with compromised support, or after secondary fixation, be cautious with a multifocal lens.
- Document the position of the IOL after fixation, including tilt in the vertical axis.
Frequently asked questions
Is there a preferred secondary fixation technique?
According to the 2020 AAO review, the data do not show clear superiority of any one technique. The choice is determined by the status of the endothelium and the angle, the integrity of the iris, the condition of the vitreous and retina, the type of existing IOL and the surgeon's experience.
Should a multifocal lens be implanted with secondary fixation?
Generally not. Multifocal lenses are particularly sensitive to tilt and decentration, and secondary fixation, even when successful, may leave a few degrees of tilt. It is worth checking the position of the lens with AS-OCT.
References
- Shen JF, Deng S, Hammersmith KM, et al. Intraocular lens implantation in the absence of zonular support: an outcomes and safety update: a report by the American Academy of Ophthalmology. Ophthalmology. 2020;127(9):1234-1258. Link
- Yamane S, Sato S, Maruyama-Inoue M, Kadonosono K. Flanged intrascleral intraocular lens fixation with double-needle technique. Ophthalmology. 2017;124(8):1136-1142. Link
- Gabriel M, Mester A, Mayer-Xanthaki C. Risk and protective factors of late in-the-bag intraocular lens dislocations: systematic review. J Cataract Refract Surg. 2025;51(1):72-83. Link
- Kimura S, Morizane Y, Shiode Y, et al. Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography. PLoS One. 2017;12(9):e0184066. Link
- Wang L, Guimaraes de Souza R, Weikert MP, Koch DD. Evaluation of crystalline lens and intraocular lens tilt using a swept-source optical coherence tomography biometer. J Cataract Refract Surg. 2019;45(1):35-40. Link
- Madrid-Costa D, Ruiz-Alcocer J, Pérez-Vives C, Ferrer-Blasco T, López-Gil N, Montés-Micó R. Visual simulation through different intraocular lenses using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg. 2012;38(6):947-958. Link
- American Academy of Ophthalmology. Basic and Clinical Science Course, Section 11: Lens and Cataract. San Francisco: AAO; 2022-2023.