In brief
- CASIA2 (Tomey): SS-OCT at a wavelength of 1310 nm, a 0.3-second scan and a 16×16 mm field. A single scan shows the anterior and posterior cornea, the angle and the entire lens.
- Tomography: detects mild keratoconus, thinning and high aberrations, and measures the posterior cornea, which is critical after refractive surgery.
- Measures the tilt and decentration of the natural lens before surgery, and of the IOL afterward.
- Grades posterior polar cataract by the state of the posterior capsule, and so estimates the risk of rupture.
- After surgery: checking the axis of the toric lens in the eye.
Why be precise? From “more or less” to measurement
20 years ago, a patient with cataract expected two things: to see better, and not to suffer a complication. Today, patients, and especially those who had refractive surgery when they were young, expect perfect vision and a life without glasses. An accurate spherical equivalent is not the whole of vision: astigmatism, night vision, contrast and peripheral vision determine the outcome just as much. And premium lenses, which were designed to meet these needs, are especially sensitive to small errors. Hence the motto of the lecture: stop estimating, and start measuring.
The device in brief
CASIA2 is a swept-source Fourier-domain OCT of the anterior segment. The fast scan (0.3 seconds) reduces motion artifacts and is comfortable for the patient and the technician. The 1310 nm wavelength penetrates the sclera well and shows the angle and the entire lens, and the wide field covers the whole cornea. The corneal maps are built from 16 radial sections, each containing at least 800 A-scans. This yields maps of the posterior surface, wide-field tomography and aberration measurements.
Versus Pentacam: studies comparing the previous model (SS-1000) with the Pentacam HR found similar repeatability. In normal eyes (Schröder 2018, 34 eyes), the Pentacam was more repeatable for anterior surface elevation, and the CASIA for pachymetry, with good repeatability for the posterior surface as well, and it was almost unaffected by misalignment. The CASIA also measured the cornea as thinner by about 12 µm, probably because the measurement does not include the tear film, although the difference is larger than the thickness of that layer. Direct comparisons of the CASIA2 found good agreement with the Pentacam for SimK, posterior curvature and total corneal power, but not for total corneal astigmatism, for which the repeatability of the CASIA2 itself was only moderate (Cheng 2024, 100 eyes). The practical conclusion: the devices are reliable, but their values are not interchangeable. For the evaluation before cataract surgery, AS-OCT adds the lens data. An important complement is an SS-OCT biometer (IOLMaster 700), which manages to measure through advanced (Mature) cataracts and shows the fovea.
Tomography before surgery: identifying the “unusual” cornea
The keratometers built into biometers sample only a few points. The IOLMaster 500 measures 6 points in a hexagonal pattern, on a ring about 2.3–2.5 mm in diameter. The Lenstar measures 32 points on two rings, 1.65 and 2.3 mm in diameter. Neither measures the posterior cornea.
The posterior surface contributes less to the optical power, because the difference in refractive index between the cornea (about 1.376) and the anterior chamber (about 1.336) is much smaller than the difference between air (1.0) and the cornea. But there are situations in which the relationship between the anterior and posterior surfaces changes, and then the posterior surface becomes critical:
- After LASIK or PRK: the anterior surface has changed, and the posterior surface has not.
- After RK: both surfaces have flattened within a small central optical zone, with extreme variability between the center and the periphery.
- After endothelial keratoplasty, or combined with it: a tendency toward a hyperopic outcome, mainly because of a change in the curvature of the posterior surface. With DMEK the mean refractive shift is about +0.4 D, and in combined surgery a target of about −0.5 D is recommended (meta-analysis, 2023). A hyperopic shift is also expected after DSAEK, and it is customary to aim for a myopic target.
- Posterior keratoconus.

Tilt and decentration of the lens
The lens analysis protocol on the CASIA2 measures the radii of the anterior and posterior surfaces of the lens, its thickness and its equatorial diameter, as well as its tilt and decentration relative to the corneal vertex. Most formulas do not use these data, but they are clinically significant.

Why does this matter? Because the tilt of the natural lens predicts, in part, that of the IOL. A Japanese study with the CASIA2 (Kimura 2017) found a mean tilt of about 5° for the natural lens and about 4.3° for the IOL, in an inferotemporal direction relative to the topographic axis, with a mean decentration of less than 0.12 mm. In 17 eyes scanned before and after surgery, the correlation was good for tilt (ICC 0.73–0.91) and moderate for decentration (ICC 0.41–0.56). With an SS-OCT biometer (Wang 2019, 333 patients), IOL tilt was on average 1.2° greater than that of the natural lens and significantly correlated with it, and tilt was greater in short eyes and with a large angle alpha. An adaptive optics simulation study (Madrid-Costa 2012, 10 eyes) showed that aspheric lenses are more sensitive than spherical lenses to decentration and tilt: a decentration of as little as 0.2 mm slightly reduced visual acuity, and at low contrast the decrease was mainly with the aspheric lenses. The differences are small, and an aspheric lens remains preferable as long as decentration is less than 0.4 mm and tilt less than 7° (Holladay).
The practical conclusion: when the natural lens is tilted or decentered, it is worth taking this into account. With a multifocal lens this is critical. With a monofocal lens it can help in choosing which lens to implant, and a spherical lens is more forgiving off-axis. A case of a scleral-sutured IOL with vertical tilt on AS-OCT is presented on the page Secondary IOL Fixation.
Astigmatism and toric lenses

Fourier analysis separates regular astigmatism, which can be corrected with a toric lens, from asymmetry and irregularity, which a toric lens will not correct. After surgery, AS-OCT also makes it possible to check the axis of the toric lens in the eye.
Posterior polar cataract: grading the risk of posterior capsule rupture
In posterior polar cataract the risk of posterior capsule rupture (PCR) is high. A study presented in the lecture (Chan et al., 2014, 37 eyes) graded the eyes by AS-OCT:
| Grade | Finding on AS-OCT | PCR in the series |
|---|---|---|
| 1 | A gap of 50% or more between the polar opacity and the posterior capsule | 1 of 19 (about 5%) |
| 2 | A gap of less than 50% | 4 of 13 (about 31%) |
| 3 | No intact posterior capsule behind the plaque, or the capsule cannot be identified (a pre-existing defect is possible) | 3 of 5 (60%) |
Overall, PCR occurred in 21.6% of eyes, similar to the literature. Comparing grade 1 with grades 2–3 (7 of 18, about 39%), sensitivity was 87.5% and specificity 62.1%. The negative predictive value was high: grade 1 eyes had no rupture in about 95% of cases. Limitations: a small retrospective study, with different surgeons. A later prospective study (Pavan Kumar 2018, 64 eyes, a single surgeon masked to the test result) graded the capsule as “intact” or “defective,” and found a sensitivity of 100%, a specificity of about 95% and a negative predictive value of 100%. High-grade cases are planned accordingly, for example hydrodelineation only, without hydrodissection, and the patient is counseled in advance about the risk.
Key points
- In posterior polar cataract, grade the risk with AS-OCT before surgery.
- After refractive surgery, RK or DMEK: tomography that includes the posterior cornea. Barrett True-K and OCT-based formulas help, but accuracy is still lower than in a normal eye.
- In advanced (Mature) cataract, try an SS-OCT biometer first, rather than contact ultrasound.
- Tilted natural lens: consider the type of IOL. With a multifocal lens this is critical.
- Check the position of the toric lens with AS-OCT.
Frequently asked questions
When should a patient be sent for tomography before cataract surgery?
After refractive surgery or RK, after corneal transplantation or DMEK, when keratoconus is suspected, when there is unexpected astigmatism or a difference in K between the eyes, and before toric or multifocal lenses.
Pentacam or AS-OCT?
Both technologies are reliable, with similar repeatability, but their values are not identical, so do not switch devices in the middle of follow-up. AS-OCT is less sensitive to misalignment, is better suited when there are corneal opacities, and adds information about the lens: thickness, tilt and decentration.
References
- Kimura S, Morizane Y, Shiode Y, et al. Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography. PLoS One. 2017;12(9):e0184066. Link
- Wang L, Guimaraes de Souza R, Weikert MP, Koch DD. Evaluation of crystalline lens and intraocular lens tilt using a swept-source optical coherence tomography biometer. J Cataract Refract Surg. 2019;45(1):35-40. Link
- Chan TC, Li EY, Yau J. Application of anterior segment optical coherence tomography to identify eyes with posterior polar cataract at high risk for posterior capsule rupture. J Cataract Refract Surg. 2014;40(12):2076-2081. Link
- Pavan Kumar G, Krishnamurthy P, Nath M, Baskaran P, Janani M, Venkatesh R. Can preoperative anterior segment optical coherence tomography predict posterior capsule rupture during phacoemulsification in patients with posterior polar cataract? J Cataract Refract Surg. 2018;44(12):1441-1445. Link
- Schröder S, Mäurer S, Eppig T, Seitz B, Rubly K, Langenbucher A. Comparison of corneal tomography: repeatability, precision, misalignment, mean elevation, and mean pachymetry. Curr Eye Res. 2018;43(6):709-716. Link
- Cheng S, Zhang J, Zhou K, Li T, Lin J, Yu AY. Repeatability and agreement of two swept-source optical coherence tomographers and Scheimpflug imaging for measurements of corneal curvature. BMC Ophthalmol. 2024;24:439. Link
- Augustin VA, Son HS, Yildirim TM, et al. Refractive outcomes after DMEK: meta-analysis. J Cataract Refract Surg. 2023;49(9):982-987. Link
- Madrid-Costa D, Ruiz-Alcocer J, Pérez-Vives C, Ferrer-Blasco T, López-Gil N, Montés-Micó R. Visual simulation through different intraocular lenses using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg. 2012;38(6):947-958. Link