In brief
- The vault, the space between the posterior surface of the ICL and the crystalline lens, is the key safety metric of this phakic lens, and it is determined mainly by the choice of lens length (12.1, 12.6, 13.2 or 13.7 mm) relative to the anatomy of the eye.
- The accepted target range is 250–750 microns; the manufacturer specifies 250–900 microns. Vault is dynamic: it decreases in bright light and with accommodation, and declines slowly over the years.
- A vault that is too low raises the risk of anterior subcapsular cataract and of lens rotation; a vault that is too high raises the risk of pressure elevation, angle closure, pigment dispersion and endothelial cell loss.
- The sizing methods: a WTW and ACD nomogram, sulcus-to-sulcus measurement with UBM, and formulas based on anterior segment OCT (NK, KS) or machine learning. All of them have a prediction error of hundreds of microns in some eyes.
- The central port of the current design makes iridotomy unnecessary, and it appears to make a low vault more tolerable; correction is possible by rotating or exchanging the lens.
What vault is and why it is dynamic
The ICL sits in the posterior chamber with its haptics resting in the sulcus, and it is designed to arch forward over the crystalline lens. Vault is measured centrally, with anterior segment OCT or UBM, between the posterior surface of the ICL and the anterior lens capsule. For a general explanation of the lens and the surgery, see The ICL Lens Guide.
Vault is a dynamic measure. In a study that measured 91 eyes with dynamic OCT, mean vault was about 476 microns in bright light compared with about 521 microns in the dark, and it also decreased with accommodation. In other words, measurement under different lighting conditions can give different results, and the conditions should be documented. Vault also declines slowly over the years: in a 10-year follow-up study (2026), a model was developed that predicts the rate of decline, and in the same series endothelial cell loss was 6.8% over the decade, with no pigment dispersion and no pupillary block.
The target range and what happens outside it

| Low vault (lens too short) | High vault (lens too long) | |
|---|---|---|
| Risks | Anterior subcapsular cataract; displacement or rotation of the lens, especially important with a toric lens | Elevated intraocular pressure, angle closure or pupillary block, a wide pupil, iris chafing and pigment dispersion, endothelial cell loss |
| Note | With the central port the risk appears to be lower: in a series of 24 eyes with a vault below 100 microns for more than 4 years, anterior opacity was found in only one eye; and in a 10-year follow-up, no ICL-induced cataract was found in the low-vault group either | In a series of 10,258 eyes, high vault was one of the two main reasons for reoperation, but in only 0.10% of eyes |
Important to remember: an out-of-range vault is not a complication in itself, but a risk factor for complications. The decision whether to intervene depends on the findings (pressure, angle, pupil, crystalline lens) and not on the number alone.
Methods for choosing lens length
| Method | What is measured | What is known |
|---|---|---|
| Manufacturer's nomogram | White-to-white (WTW) and anterior chamber depth (ACD) | The traditional method. WTW is an indirect measure of sulcus size; manual measurement under the microscope, with the patient supine, improves it. In a study comparing two nomograms and four formulas (214 eyes), WTW from the Pentacam with the manufacturer's nomogram performed well, and with all methods only about half of the eyes, or fewer, received the “ideal” size. |
| UBM | Sulcus-to-sulcus diameter (STS) and the position of the crystalline lens relative to it | Directly measures where the haptics rest. In one nomogram (72 eyes), mean vault was about 340 microns, and the size chosen differed from that of the WTW method in 69% of eyes. |
| NK formula | Anterior segment OCT: the distance between the scleral spurs (ACW) and the crystalline lens rise (CLR) | Developed in 2018 and refined several times; with the second version, about 91% of eyes fell within the 250–1,000 micron range. Built into the OCT device on which it was developed. |
| KS formula | OCT: angle-to-angle distance (ATA) | A simple linear formula of lens length minus ATA; it too is built into the same device. |
| Machine learning | A combination of many parameters from OCT devices | In a series of 1,745 eyes, the prediction error was smaller than with the standard nomogram, but its performance differed between validation sets from two countries; in other words, formulas need to be calibrated to the population. |
How accurate are the formulas? In a head-to-head comparison (2024, 106 eyes), both overestimated the vault by 70–100 microns on average, and the limits of agreement were about ±500 microns. In another study, the mean error was small with the KS formula and larger with the NK formula, and both tended to overestimate when a large lens was chosen. The anatomy of the ciliary body also plays a part: a narrow iris-ciliary angle and an anteriorly positioned ciliary body were associated with a risk of vault above 1,000 microns. The practical conclusion: no formula removes the need to measure the vault after surgery.
A practical approach to choosing lens length
- Measure WTW with at least two devices, as well as ACD (from the endothelium) and CLR; check that the measurements agree.
- When the WTW is borderline between two sizes, when it does not match the internal dimensions, or when one of the extreme sizes is involved (12.1 or 13.7 mm), add a direct measurement of STS with UBM or of ATA with OCT. According to a 2024 review, the formulas are less accurate precisely at the extreme sizes.
- Compare at least two methods; when they disagree, repeat the measurements and consider direct measurement of the internal dimensions.
- After surgery: measure vault, pressure and angle configuration at the early follow-up visits and thereafter, and document the lighting conditions.
When the vault or the axis is not as planned
In a series of 10,258 eyes with the central-port lens (2022), only 0.21% required reoperation. The two main reasons were axis misalignment of a toric lens (0.12%; in most cases treated by re-rotation, with mean residual cylinder falling from −1.75 D to −0.87 D), and high vault (0.10%; treated by exchange for a smaller lens or by rotating the lens to the vertical axis). According to the authors, rotation to the vertical axis is a less invasive alternative in some cases of high vault. In other series it was low vault that was the leading reason for exchange, a reminder that outcomes depend on the population and on the sizing method.
On the central port: the current design has a 0.36 mm port in the optic, which allows aqueous flow from the posterior chamber to the anterior chamber and makes preoperative iridotomy unnecessary. In a randomized fellow-eye study (32 patients, 5 years), the visual and safety outcomes were similar to those of the lens without a port, and a single anterior subcapsular cataract appeared only in an eye with the older lens.
Bottom line for the surgeon
- The choice of lens length is the most important decision after the indication itself; base it on at least two methods.
- Direct measurement of the internal dimensions (STS or ATA) adds information mainly in borderline cases and at the extreme sizes.
- Vault is a dynamic number: document the conditions, and decide on intervention according to the clinical findings and not the number alone.
- High vault with elevated pressure or a narrow angle: exchange or vertical rotation; a toric axis that has shifted: re-rotation.
- Before ICL in a borderline cornea, see Ectasia Screening.
Frequently asked questions
What is the ideal vault after ICL?
The range accepted as the target is 250–750 microns; the manufacturer specifies 250–900 microns, or 50%–180% of corneal thickness. Vault changes with light, with accommodation and over the years, so the measurement conditions are documented and decisions are made according to the clinical findings.
Does a low vault require lens exchange?
Not necessarily. With central-port lenses, long-term series have found very few opacities even with low vault. The decision is based on the state of the crystalline lens, axis stability with a toric lens and follow-up over time.
Are the OCT formulas better than WTW?
They add information about the internal dimensions, but in head-to-head comparisons their prediction error can still reach hundreds of microns, and in one of the studies WTW with the manufacturer's nomogram performed well. It is worth comparing at least two methods, especially in borderline cases.
References
- Thompson V, Cummings AB, Wang X. Implantable collamer lens procedure planning: a review of global approaches. Clin Ophthalmol. 2024;18:1033-1043. Link
- Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Implantable collamer lens sizing method based on swept-source anterior segment optical coherence tomography. Am J Ophthalmol. 2018;187:99-107. Link
- Tang C, Sun T, Duan H, Liu Y, Qi H. Evaluation of the performance of two nomograms and four vault prediction formulas for implantable collamer lens size selection. J Refract Surg. 2023;39(7):456-461. Link
- Zhong X, Li Y, Li Y, Wang G, Du Y, Zhang M. Comparison of predictability in vault using NK formula and KS formula for the implantable collamer lens surgery. J Ophthalmol. 2024;2024:4256371. Link
- Wei R, Li M, Aruma A, et al. Factors leading to realignment or exchange after implantable collamer lens implantation in 10 258 eyes. J Cataract Refract Surg. 2022;48(10):1190-1196. Link
- Alfonso-Bartolozzi B, Fernández-Vega-Cueto L, Lisa C, et al. Ten-year follow-up of posterior chamber phakic intraocular lens with central port design in patients with low and normal vault. J Cataract Refract Surg. 2024;50(5):441-447. Link
- Kato S, Shimizu K, Igarashi A. Vault changes caused by light-induced pupil constriction and accommodation in eyes with an implantable collamer lens. Cornea. 2019;38(2):217-220. Link
- Nakamura T, Isogai N, Kojima T, Sugiyama Y, Yoshida Y. Vault decay prediction formula for posterior chamber phakic intraocular lens with central hole implantation: a 10-year follow-up study. Am J Ophthalmol. 2026;282:49-59. Link